英文摘要To minimize the uncertainties related to the FTIR spectrometer, three improved methods were proposed to measure the spectral emissivity. For method 1 (M-1) and method 2 (M-2), an invented device was used to change the environment radiation which makes the ill conditioned equation solvable. Method 3 (M-3) is based on the gray body assumption, and an iterative procedure was applied to derive the temperature of samples, then the spectral emissivity can be retrieved based on this temperature and the FTIR measurements. In order to validate these methods, the emissivity derived from the measured temperature was used as a reference. The results indicate that for the samples with high emissivity, such as quartz plates and cardboard, three improved methods have comparable accuracy; while for the samples with low emissivity, such as rusty iron plate and aluminum plate, the M-3 is more reliable than the other two methods. M-1 and M-2 need to go through more procedures than M-3, and this may lead to some deviation from the reality. Overall, three improved methods greatly improved the spectral emissivity measurements with FTIR spectrometer.
出处Journal of Infrared and Millimeter Waves
32
4
366-371
收录类别SCI
语种英语
ISSN号1001-9014
内容类型SCI/SSCI论文
源URL[http://ir.igsnrr.ac.cn/handle/311030/30098]  
专题地理科学与资源研究所_历年回溯文献
推荐引用方式
GB/T 7714
Tian J.. Three improved methods for measuring spectral emissivity based on the FTIR spectrometer. 2013.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace