Characteristics of a prototype CdZnTe detector
Xu, HC(徐慧超) ; Cheng, C ; Zhao, CL(赵翠兰) ; Zhang, JZ(张金洲) ; Pu, SJ
刊名NUCLEAR SCIENCE AND TECHNIQUES
2007
卷号18期号:6页码:358
ISSN号1001-8042
通讯作者Xu, HC (reprint author), Chinese Acad Sci, Shanghai Inst Appl Phys, Shanghai 201800, Peoples R China
英文摘要Cadmium zinc telluride (CZT) is a preferred material for X-ray and gamma-ray detector. Thanks to the relatively high atomic number, high density and wide band-gap, CZT detector possesses sharp energy resolution and high detection efficiency without cryogenic cooling. We have developed a CdZnTe detector with an energy resolution of 3.45% (FWHM) at 59.54 keV at room temperature, and it is used for X-ray fluorescence analysis. In this paper, leakage current, energy resolution and long term stability of the CZT detector are discussed.
学科主题Chemistry; Electrochemistry; Instruments & Instrumentation
收录类别SCI
语种英语
WOS记录号WOS:000255053200008
公开日期2012-04-18
内容类型期刊论文
源URL[http://ir.sinap.ac.cn/handle/331007/7672]  
专题上海应用物理研究所_中科院上海应用物理研究所2004-2010年
推荐引用方式
GB/T 7714
Xu, HC,Cheng, C,Zhao, CL,et al. Characteristics of a prototype CdZnTe detector[J]. NUCLEAR SCIENCE AND TECHNIQUES,2007,18(6):358.
APA Xu, HC,Cheng, C,Zhao, CL,Zhang, JZ,&Pu, SJ.(2007).Characteristics of a prototype CdZnTe detector.NUCLEAR SCIENCE AND TECHNIQUES,18(6),358.
MLA Xu, HC,et al."Characteristics of a prototype CdZnTe detector".NUCLEAR SCIENCE AND TECHNIQUES 18.6(2007):358.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace