CORC  > 北京大学  > 信息科学技术学院
A novel test method of low leakage rate of MEMS vacuum packaging
Li, Lei ; Li, Jie ; Zhao, Zhao ; Wang, Han ; Zhang, Wei ; Kan, Jinsong
2015
英文摘要With the development of MEMS technology, especially the development of packaging technology, more and more MEMS devices require the vacuum packaging to improve their performances. However, MEMS vacuum packaging technology has been a technical problem for many years. Currently, we still can't truly achieve the low leakage rate vacuum packaging of MEMS devices. Meanwhile, the existing detecting method can't accurately measure low leakage rate. For these reasons, this paper carried out a novel test method to detect the low leakage rate of MEMS vacuum packaging. The use of a crystal to real-time monitoring the vacuum level inside of the envelope to get indirect leakage rate of the envelope was proposed. Through the work of crystal calibration, data testing, modification and re-testing, finally, we got a MEMS vacuum packaging envelope with the leakage rate of 3.2??10-11Pa??m3/s by the FC135 Patch crystal. ? 2015 IEEE.; EI; 175-178
语种英语
出处10th IEEE International Conference on Nano/Micro Engineered and Molecular Systems, NEMS 2015
DOI标识10.1109/NEMS.2015.7147404
内容类型其他
源URL[http://ir.pku.edu.cn/handle/20.500.11897/423615]  
专题信息科学技术学院
推荐引用方式
GB/T 7714
Li, Lei,Li, Jie,Zhao, Zhao,et al. A novel test method of low leakage rate of MEMS vacuum packaging. 2015-01-01.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace