CORC  > 北京大学  > 信息科学技术学院
High-efficiency blood separation utilizing spiral filtration microchannel with gradually varied width
Ju, Y.R. ; Geng, Z.X. ; Zhang, L.Q. ; Wang, W. ; Li, Z.H.
2011
英文摘要A hybrid microfluidic separation device that uses the crossflow and the centrifugation effect to separate human plasma from blood cells has been designed and fabricated. The chip mainly consists of the spiral microchannel, which is divided into inner and outer mcirochannels by micropillar arrays. The slits between the micropillars act as filter. Clogging and jamming in this filtration structure are efficiently alleviated. Remarkably, the width of the inner microchannel is gradually decreasing from inlet to outlet in order to increase separation efficiency. The performances of the separation device were investigated theoretically and experimentally. Due to high separation efficiency and compact structure, it is envisaged that this device can be integrated with other microfluidic device for point-of-care diagnostics in the near future. ? 2011 IEEE.; EI; 0
语种英语
DOI标识10.1109/TRANSDUCERS.2011.5969447
内容类型其他
源URL[http://ir.pku.edu.cn/handle/20.500.11897/295212]  
专题信息科学技术学院
推荐引用方式
GB/T 7714
Ju, Y.R.,Geng, Z.X.,Zhang, L.Q.,et al. High-efficiency blood separation utilizing spiral filtration microchannel with gradually varied width. 2011-01-01.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace