大气中挥发性有机物在线监测系统; Online monitoring system for volatile organic compounds in the atmosphere | |
刘兴隆 ; 曾立民 ; 陆思华 ; 于雪娜 | |
刊名 | 环境科学学报 |
2009 | |
关键词 | 挥发性有机化合物 低温捕集 在线监测 气相色谱-质谱仪 VOCs cryn-trapping online monitoring GC-FID/MS |
英文摘要 | 一种新型的监测大气中挥发性有机物(包括含氧挥发性有机物)的在线监测系统被研制,即将超低温冷阱捕集-热解析装置与气相色谱-质谱仪联用.其分析方法是大气样品经除水、除O_3后以60 mL·min~(-1)的流速通过温度为-150℃的超低温冷阱捕集5min,然后样品在110℃下解析后进入GC-FID/MS系统进行分析,时间分辨率为1h.系统使用混和标气进行标定.目标化合物定量曲线的R~2值为0.9137~0.9998,相对标准偏差(RSD)均小于10%.将系统与相关商业化的VOC在线监测仪器进行比对,对于相同目标化合物进行分析,其相关系数r在0.7412~0.9620之间.; A new online monitoring system for atmospheric volatile organic compounds, including oxygenated volatile organic compounds, was developed with a combination of cryo-trapping-thermal desorption and GC-FID/MS ( Gas chromatography-Flame Ionization Detector/Mass Spectrometry). Water and ozone were removed to avoid interference. Ambient air was collected with an air flow rate of 60 mL. min~(-1) at-15℃ for S minutes, then the sample was desorbed by beating to 110℃ and the volatile organic compounds were detected by GC-FID/MS. The time resolution is 1 h. The system was calibrated with multi-component standards. The calibration results indicated good regression results for target compounds with R~2 = 0. 9137 -0. 9998 and relative standard deviation ( BSD% )below 10%. The newly developed system showed a similar variation of VOC species concentrations to those currently used incommercial online instruments, and the correlation coefficients (r) ranged between 0. 7412 and 0. 9620.; 国家高技术研究发展计划(863计划); 中文核心期刊要目总览(PKU); 中国科技核心期刊(ISTIC); 中国科学引文数据库(CSCD); 0; 12; 2471-2477; 29 |
语种 | 中文 |
内容类型 | 期刊论文 |
源URL | [http://ir.pku.edu.cn/handle/20.500.11897/4570] |
专题 | 环境科学与工程学院 |
推荐引用方式 GB/T 7714 | 刘兴隆,曾立民,陆思华,等. 大气中挥发性有机物在线监测系统, Online monitoring system for volatile organic compounds in the atmosphere[J]. 环境科学学报,2009. |
APA | 刘兴隆,曾立民,陆思华,&于雪娜.(2009).大气中挥发性有机物在线监测系统.环境科学学报. |
MLA | 刘兴隆,et al."大气中挥发性有机物在线监测系统".环境科学学报 (2009). |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论