Selective spectrophotometric determination of human serum albumin with tetraiodo phenol sulfonphthalein
Ma, CQ ; Li, KA ; Tong, SY
刊名analytical letters
1997
关键词spectrophotometry protein tetraiodo phenol sulfonphthalein DYE-BINDING PROTEINS BLUE ASSAY
英文摘要Tetraiodo phenol sulfonphthalein (TIPSP) reacts with protein rapidly to form the dye-protein complex with an absorbance maximum at 617 nm. Water is used as the blank instead of the reagent blank because the reagents have no absorption at this wavelength; thus the operating steps are simplified. Under the experimental conditions (0.03 mM TIPSP, pH 2.63), the linear range of this assay is composed of three parts: 10.2-54.4 mu g . ml(-1), 54.4-95.2 mu g . ml(-1) and 95.2-136.0 mu g . ml(-1). Most amino acids, reducing reagents (e.g. ascorbic acid) and complexing reagents (e.g. EDTA) don't interfere with the assay. The interferences from detergents, salts and ethanol can be minimized by dilution. In detection of human serum albumin, this method gave results close to that of the bromocresol green assay.; Chemistry, Analytical; SCI(E); 4; ARTICLE; 4; 739-752; 30
语种英语
内容类型期刊论文
源URL[http://ir.pku.edu.cn/handle/20.500.11897/401928]  
专题化学与分子工程学院
推荐引用方式
GB/T 7714
Ma, CQ,Li, KA,Tong, SY. Selective spectrophotometric determination of human serum albumin with tetraiodo phenol sulfonphthalein[J]. analytical letters,1997.
APA Ma, CQ,Li, KA,&Tong, SY.(1997).Selective spectrophotometric determination of human serum albumin with tetraiodo phenol sulfonphthalein.analytical letters.
MLA Ma, CQ,et al."Selective spectrophotometric determination of human serum albumin with tetraiodo phenol sulfonphthalein".analytical letters (1997).
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace