A MICROSAMPLING SYSTEM FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY USING HYDRODYNAMIC INJECTION | |
XU, SK; FANG, ZL | |
刊名 | MICROCHEMICAL JOURNAL |
1995 | |
卷号 | 51期号:3页码:360-366 |
ISSN号 | 0026-265X |
英文摘要 | A flow injection microsample introduction system for flame atomic absorption spectrometry is developed based on air-transported sample loading and hydrodynamic injection. An augmenting water stream was used between sample loop and nebulizer to prevent disturbances from nebulizer suction during sample loading. A total of 0.8 mi of sample was sufficient for triplicate measurements, including washout between samples, and achieving the sensitivity of conventional steady-state measurements. The system was applied to the determination of copper in serum samples at a sampling frequency of 100/h and with a precision of about 2% rsd. (C) 1995 Academic Press, Inc. |
学科主题 | Chemistry, Analytical |
语种 | 英语 |
公开日期 | 2011-09-23 |
内容类型 | 期刊论文 |
源URL | [http://210.72.129.5/handle/321005/55719] |
专题 | 沈阳应用生态研究所_沈阳应用生态研究所 |
推荐引用方式 GB/T 7714 | XU, SK,FANG, ZL. A MICROSAMPLING SYSTEM FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY USING HYDRODYNAMIC INJECTION[J]. MICROCHEMICAL JOURNAL,1995,51(3):360-366. |
APA | XU, SK,&FANG, ZL.(1995).A MICROSAMPLING SYSTEM FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY USING HYDRODYNAMIC INJECTION.MICROCHEMICAL JOURNAL,51(3),360-366. |
MLA | XU, SK,et al."A MICROSAMPLING SYSTEM FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY USING HYDRODYNAMIC INJECTION".MICROCHEMICAL JOURNAL 51.3(1995):360-366. |