Low temperature testing and neutron irradiation of a swept charge device on board the HXMT satellite
Liu Yan; Zhu Yue; Lu Bo; Zhang Yi; Yang Yan-Ji; Cui Wei-Wei; Li Mao-Shun; Liu Xiao-Yan; Wang Juan; Han Da-Wei
刊名CHINESE PHYSICS C
2012
卷号36期号:10页码:991-995
关键词SCD HXMT energy resolution readout noise neutron irradiation
ISSN号1674-1137
DOI10.1088/1674-1137/36/10/013
英文摘要We present the low temperature testing of an SCD detector, investigating its performance such as readout noise, energy resolution at 5.9 keV and dark current. The SCD's performance is closely related to temperature, and the temperature range of -80 degrees C to -50 degrees C is the best choice, where the FWHM at 5.9 key is about 130 eV. The influence of the neutron irradiation from an electrostatic accelerator with fluence up to 1 x 10(9) cm(-2) has been examined. We find the SCD is not vulnerable to neutron irradiation. The detailed operations of the SCD and the test results of low temperature are reported, and the results of neutron irradiation are discussed.
WOS研究方向Physics, Nuclear ; Physics, Particles & Fields
语种英语
WOS记录号WOS:000309900800013
内容类型期刊论文
源URL[http://ir.ihep.ac.cn/handle/311005/282218]  
专题硬X射线调制望远镜卫星
推荐引用方式
GB/T 7714
Liu Yan,Zhu Yue,Lu Bo,et al. Low temperature testing and neutron irradiation of a swept charge device on board the HXMT satellite[J]. CHINESE PHYSICS C,2012,36(10):991-995.
APA Liu Yan.,Zhu Yue.,Lu Bo.,Zhang Yi.,Yang Yan-Ji.,...&Zhang Zi-Liang.(2012).Low temperature testing and neutron irradiation of a swept charge device on board the HXMT satellite.CHINESE PHYSICS C,36(10),991-995.
MLA Liu Yan,et al."Low temperature testing and neutron irradiation of a swept charge device on board the HXMT satellite".CHINESE PHYSICS C 36.10(2012):991-995.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace