Chronically Implantable Package Based on Alumina Ceramics and Titanium with High-density Feedthroughs for Medical Implants
Hangao Yang; Tianzhun Wu; Saisai Zhao; Shanshan Xiong; Bo Peng; Mark S. Humayun
2018
会议日期2018
会议地点夏威夷
英文摘要Implantable package to hermetically encapsulate electronics inside human body is critical for active implant devices such as neuroprotheses. To meet the demanding package requirement for smaller size and higher feedthrough density, we propose a high-density (100+ feedthroughs for 10 mm diameter) ceramic/metal composite package with helium leakage rate on the 10 -10 Pa×m 3 /s, at the same time possessing the best cytotoxicity level of Grade 0, which enable the chronic implant in human. Pure alumina substrate co-sintered with platinum (Pt) paste filled in micrometer holes have demonstrated extremely good hermetical seal and biocompatibility, then its braze joint with a titanium(Ti) ring was achieved, followed by the laser welding with a Ti cap. Standard helium leakage rate and cytotoxicity experiments have shown each component and joint interface are qualified for 100-year chronic implant, which is significant for various active implant instruments.
URL标识查看原文
内容类型会议论文
源URL[http://ir.siat.ac.cn:8080/handle/172644/14524]  
专题深圳先进技术研究院_医工所
推荐引用方式
GB/T 7714
Hangao Yang,Tianzhun Wu,Saisai Zhao,et al. Chronically Implantable Package Based on Alumina Ceramics and Titanium with High-density Feedthroughs for Medical Implants[C]. 见:. 夏威夷. 2018.