Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes | |
Luo, Yong; Wu, Dapeng; Zeng, Shaojiang; Gai, Hongwei; Long, Zhicheng; Shen, Zheng; Dai, Zhongpeng; Qin, Jianhua; Lin, Bingcheng | |
刊名 | analytical chemistry |
2006-09-01 | |
卷号 | 78期号:17页码:6074-6080 |
通讯作者 | 秦建华 ; 林炳承 |
产权排序 | 1;1 |
英文摘要 | a novel sample injection method for chip ce was presented. this injection method uses hydrostatic pressure, generated by emptying the sample waste reservoir, for sample loading and electrokinetic force for dispensing. the injection was performed on a double-cross microchip. one cross, created by the sample and separation channels, is used for formation of a sample plug. another cross, formed by the sample and controlling channels, is used for plug control. by varying the electric field in the controlling channel, the sample plug volume can be linearly adjusted. hydrostatic pressure takes advantage of its ease of generation on a microfluidic chip, without any electrode or external pressure pump, thus allowing a sample injection with a minimum number of electrodes. the potential of this injection method was demonstrated by a four-separation-channel chip ce system. in this system, parallel sample separation can be achieved with only two electrodes, which is otherwise impossible with conventional injection methods. hydrostatic pressure maintains the sample composition during the sample loading, allowing the injection to be free of injection bias. |
WOS标题词 | science & technology ; physical sciences |
类目[WOS] | chemistry, analytical |
研究领域[WOS] | chemistry |
关键词[WOS] | capillary electrophoresis ; microchip electrophoresis ; microfluidic chips ; hydrodynamic injection ; flow ; performance ; system ; devices ; separations |
收录类别 | SCI |
原文出处 | true |
语种 | 英语 |
WOS记录号 | WOS:000240158000024 |
公开日期 | 2010-11-30 |
内容类型 | 期刊论文 |
源URL | [http://159.226.238.44/handle/321008/97221] |
专题 | 大连化学物理研究所_中国科学院大连化学物理研究所 |
作者单位 | Chinese Acad Sci, Dalian Inst Chem Phys, Lab Microfluid Syst Biol, Dalian, Liaoning Prov, Peoples R China |
推荐引用方式 GB/T 7714 | Luo, Yong,Wu, Dapeng,Zeng, Shaojiang,et al. Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes[J]. analytical chemistry,2006,78(17):6074-6080. |
APA | Luo, Yong.,Wu, Dapeng.,Zeng, Shaojiang.,Gai, Hongwei.,Long, Zhicheng.,...&Lin, Bingcheng.(2006).Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes.analytical chemistry,78(17),6074-6080. |
MLA | Luo, Yong,et al."Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes".analytical chemistry 78.17(2006):6074-6080. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论