CORC  > 清华大学
A method for dynamically measuring the Soft Tissue Balance in the Total Knee Replacement
Wu, Han ; Wang, You ; Dai, Kerong ; Wei, Qihang ; Ye, Datian ; Li, Xiuwang ; Zhao, Ming
2010-05-10 ; 2010-05-10
会议名称WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6 ; World Congress on Medical Physics and Biomedical Engineering ; Seoul, SOUTH KOREA ; Web of Science
关键词Soft tissue balance Total knee replacement load sensor spacer Engineering, Biomedical Physics, Applied Imaging Science & Photographic Technology
中文摘要Osteoarthritis, is one of the most common arthropathy in the world, and the artificial knee replacement surgery becomes one of the major measures to cure the severe pathological changes in the knee. The soft tissue balance is very important to prolong the artificial joint's life in the total knee replacement (TKR), but now the balance is mainly depended on the surgeon's experience, so the result is not confident. In order to meet the quantitative soft tissues balance measurement requirement in the TKR, a new soft tissues balance measurement method and system have been developed based on the load sensor technology, computer graphics technology and the precise mechanical design in this paper. In the system, six precise load sensors have been embedded in the front end measure device, which will be inserted into the knee joint. Each sensor is responsive to a force generated between the condyle and the device, and the pressure of the sensors represents the force of the posterior, middle, and anterior position in the medial/lateral part of the knee joint. Furthermore, the device has the function that can change the distance between its upper and lower plate by the mechanical structure for adjusting the space in the knee joint. The preliminary test shows that the measurement system can accomplish the force measurement and the distance adjustment in the knee joint, and display the data of the force distribution in three-dimensional manner in real-time. Otherwise, all the reliable force and distance data during measuring saved can be provided for the modeling of the knee joint dynamics.
会议录出版者SPRINGER-VERLAG BERLIN ; BERLIN ; HEIDELBERGER PLATZ 3, D-14197 BERLIN, GERMANY
语种英语 ; 英语
内容类型会议论文
源URL[http://hdl.handle.net/123456789/19941]  
专题清华大学
推荐引用方式
GB/T 7714
Wu, Han,Wang, You,Dai, Kerong,et al. A method for dynamically measuring the Soft Tissue Balance in the Total Knee Replacement[C]. 见:WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, World Congress on Medical Physics and Biomedical Engineering, Seoul, SOUTH KOREA, Web of Science.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace