便携式体外生命支持系统离体和在体实验研究 | |
刘宇; 周圆圆![]() ![]() | |
刊名 | 中国体外循环杂志
![]() |
2017 | |
卷号 | 15期号:1页码:38-41 |
关键词 | 便携 体外生命支持 动物实验 体外实验 |
ISSN号 | 1672-1403 |
其他题名 | Portable extracorporeal life support system in vitro and in vivo experiment |
通讯作者 | 王辉山 |
产权排序 | 2 |
中文摘要 | 目的初步考察便携式体外生命支持系统的离体和在体性能。方法离体实验检测便携式体外生命支持系统的稳定性;在体实验中12只雌性巴马香猪随机分为实验组和对照组,每组6只。分别应用便携式体外生命支持系统及Maquet公司Rotaflow系统进行体外循环,4.0 L/min运转6 h,于转流前、转流中每2 h及转流后取静脉血检测游离血红蛋白含量评价装置溶血性能并检测血小板激活率。结果离体实验证实便携式体外生命支持系统稳定性优于或等于同类Maquet公司Rotaflow系统;在体实验提示4.0 L/min流量运行6 h,血浆游离血红蛋白含量均在50 g/L以下,血小板激活均小于8%。结论便携式体外生命支持系... |
英文摘要 | Objective: To explore the performance of a portable extracorporeal life support (PELS) system in vitro and in vivo. Methods: In vitro,PELS system was tested for stable assessment; in vivo,12 female swine were randomized into experiment group (n =6) and control group (n=6). Cardiopulmonary bypass (CPB) was respectively established with PELS and RotaFlow with 4 L/min of flow for 6 h. The plasma free hemoglobin and platelet activation were evaluated before initial of CPB,every 2 hs in CPB and after terminated CPB. Results In vitro,experiments confirmed that the stability of PELS system was better than or equal to Maquet company Ro taflow system; in vivo,the plasma free hemoglobin was lower than 20 g/L and the platelet activation was lower than 10% after 6 h performance of PELS system. Conclusion: The portable extracorporeal life support system conducts the reliable performance in vitro and in vivo. |
语种 | 中文 |
内容类型 | 期刊论文 |
源URL | [http://ir.sia.cn/handle/173321/20371] ![]() |
专题 | 沈阳自动化研究所_机器人学研究室 |
推荐引用方式 GB/T 7714 | 刘宇,周圆圆,尹宗涛,等. 便携式体外生命支持系统离体和在体实验研究[J]. 中国体外循环杂志,2017,15(1):38-41. |
APA | 刘宇,周圆圆,尹宗涛,李洪谊,&王辉山.(2017).便携式体外生命支持系统离体和在体实验研究.中国体外循环杂志,15(1),38-41. |
MLA | 刘宇,et al."便携式体外生命支持系统离体和在体实验研究".中国体外循环杂志 15.1(2017):38-41. |
个性服务 |
查看访问统计 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论