A Full Custom Analog Front-End for Long-time ECG Monitoring | |
Wen, Meiying; Cheng, Yayu; Li, Ye | |
2013 | |
会议名称 | 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013 |
会议地点 | Osaka, Japan |
英文摘要 | An analog front-end (AFE) used in portable electrocardiogram (ECG) monitoring devices is proposed. This AFE has included all necessary functions for the commercial applications. The core circuit consists of the instrumentation amplifier (IA), a 2nd order Butterworth low pass filter, and the second amplifying stage. The driven-right-leg circuit is integrated in the IA to effectively suppress the common mode interference. And the power management circuits provide a stable supply voltage, bias current and reference voltage for the other circuits. To guarantee the validity of the continuous monitoring data, the leadoff monitoring circuit is developed to monitor the connection of the leads. The chip is taped out with SMIC 0.18μm CMOS process, and the measured results show that the common mode rejection ratio (CMRR) and power supply rejection ratio (PSRR) achieve 75dB and 90dB respectively, and the equivalent input referred noise is 12μV. |
收录类别 | EI |
语种 | 英语 |
内容类型 | 会议论文 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/4970] |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2013 |
推荐引用方式 GB/T 7714 | Wen, Meiying,Cheng, Yayu,Li, Ye. A Full Custom Analog Front-End for Long-time ECG Monitoring[C]. 见:2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013. Osaka, Japan. |