通讯作者Fu Yanyan, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China. ; Li Dawei, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China. ; Liu Xiaofeng, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China. ; Zhao Yuanan, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China. ; Guo Meng, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China. ; Zhang Lei, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Science, Shanghai 201800, China.
中文摘要基于表面热透镜技术搭建了光学薄膜弱吸收测量系统,使用不同功率的抽运光辐照样品,并通过光束质量分析仪记录经样品表面反射后的探测光强度的空间分布,以探测光强度最强点所对应的像素坐标为中心两边对称取多个像素点表示不同尺寸的接收器,从而分析不同尺寸光电信号接收器得到的光热信号强度。结果表明,随着接收器尺寸的增加,探测面积覆盖范围超过探测光束峰值区域,得到的光热信号与抽运光功率的比值逐渐偏离线性,从而引入的测量误差也就增大。因此,实际测量中应在保证测量极限的前提下尽量缩小探测器尺寸以提高测量的准确度。
英文摘要Based on the weak absorption measurement system with surface thermal lens (STL) technique, different power pump laser is used to irradiate the sample and record the intensity distribution of probe beam with beam quality analyzer. Making the pixel coordinates of the strongest intensity point of the probe beam as the center, by taking multiple pixel from both sides of the center, different sizes of photoelectric signal receiver can be represented and photo- thermal signal intensity which obtained by different sizes of photoelectric signal receiver can be analyzed. The photo-thermal signal as a function of size of photoelectric signal receiver is experimental measured. The experimental results show that with the size of photoelectric signal receiver increasing, the coverage area of detection exceeds the probe beam peak area,and then the photo-thermal signal deviates from linearity gradually, which enlarges the measurement error. Therefore, to improve the accuracy of measurements,the actual size of photoelectric signal receiver should be minimized on the premise of guaranteeing measurement limit.
收录类别EI
原文出处http://www.opticsjournal.net
语种中文
内容类型期刊论文
版本出版稿
源URL[http://ir.siom.ac.cn/handle/181231/13839]  
专题上海光学精密机械研究所_中科院强激光材料重点实验室
作者单位1.符燕燕, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
2.李大伟, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
3.刘晓凤, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
4.赵元安, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
5.郭猛, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
6.张蕾, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
推荐引用方式
GB/T 7714
符燕燕,李大伟,刘晓凤,等. 光电信号接收器尺寸对光学薄膜弱吸收测量的影响[J]. 中国激光,2015,42(7):707001.
APA 符燕燕,李大伟,刘晓凤,赵元安,郭猛,&张蕾.(2015).光电信号接收器尺寸对光学薄膜弱吸收测量的影响.中国激光,42(7),707001.
MLA 符燕燕,et al."光电信号接收器尺寸对光学薄膜弱吸收测量的影响".中国激光 42.7(2015):707001.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace