正弦相位调制全深度频域多普勒光学相干层析成像技术
南楠; 步鹏; 李中梁; 郭昕; 王向朝
刊名中国激光
2014
卷号41期号:2页码:204001
关键词医用光学 医疗与生物成像 光学相干层析 正弦相位调制 多普勒探测
其他题名Full-Range Fourier Domain Doppler Optical Coherence Tomography Based on Sinusoidal Phase Modulation
中文摘要将多普勒探测与正弦相位调制复频域光学相干层析成像技术(OCT)相结合,建立了基于正弦相位调制的全深度频域多普勒光学相干层析成像系统。利用正弦相位调制B-M扫描方法,并采用傅里叶变换结合带通滤波的方法重建复干涉谱信号,获得全深度层析及多普勒图像。成像深度范围扩大为原来的2倍,在整幅图像范围内均可获得较高的速度探测灵敏度。基于该系统获得了血流仿体的全深度层析图及全深度多普勒图。实验测量的系统最小可探测速度为5. 35 mum/s。
英文摘要A full-range Fourier domain Doppler optical coherence tomography (DOCT) system based on sinusoidal phase-modulating method is developed. The system combines Doppler detection with complex Fourier domain OCT. Full-range OCT and Doppler images are achieved using the complex spectral interferogram, which is retrieved based on the sinusoidal phase modulation B-M scanning and the combination of Fourier transform analysis and bandpass filter method. With the system, the depth imaging range is doubled and high velocity sensitivity is available in the whole Bscan image. Full-range OCT and full-range Doppler images of a flow phantom are achieved with the proposed system. The minimum detectable velocity of the system is investigated, which is 5.35 mum/s.
收录类别EI
语种中文
内容类型期刊论文
版本出版稿
源URL[http://ir.siom.ac.cn/handle/181231/13676]  
专题上海光学精密机械研究所_信息光学与光电技术实验室
作者单位1.南楠, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
2.李中梁, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
3.郭昕, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
4.王向朝, 中国科学院上海光学精密机械研究所, 上海 201800, 中国.
5.步鹏, 卡尔蔡司光学(上海)有限公司, 上海 200127, 中国.
推荐引用方式
GB/T 7714
南楠,步鹏,李中梁,等. 正弦相位调制全深度频域多普勒光学相干层析成像技术[J]. 中国激光,2014,41(2):204001.
APA 南楠,步鹏,李中梁,郭昕,&王向朝.(2014).正弦相位调制全深度频域多普勒光学相干层析成像技术.中国激光,41(2),204001.
MLA 南楠,et al."正弦相位调制全深度频域多普勒光学相干层析成像技术".中国激光 41.2(2014):204001.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace