Quantitative analysis of single aerosol particles using polycapillary X-ray optics | |
Sun, TX; Liu, ZG; Li, YD; Wang, GF; Zhu, GH; Ding, XL; Xu, Q; Liu, H; Luo, P; Pan, QL | |
刊名 | SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY |
2009 | |
卷号 | 64期号:11-12页码:1194-1197 |
关键词 | Micro X-ray fluorescence Capillary X-ray optics Single aerosol particle analysis |
通讯作者 | [Sun, Tianxi ; Liu, Zhiguo ; Li, Yude ; Wang, Guangfu ; Zhu, Guanghua ; Ding, Xunliang ; Liu, Hui ; Luo, Ping ; Pan, Qiuli ; Lin, Xiaoyan ; Teng, Yuepeng] Beijing Normal Univ, Coll Nucl Sci & Technol, Beijing 100875, Peoples R China ; [Sun, Tianxi ; Liu, Zhiguo ; Li, Yude ; Wang, Guangfu ; Zhu, Guanghua ; Ding, Xunliang ; Liu, Hui ; Luo, Ping ; Pan, Qiuli ; Lin, Xiaoyan ; Teng, Yuepeng] Beijing Normal Univ, Key Lab Beam Technol & Mat Modificat, Minist Educ, Beijing 100875, Peoples R China ; [Sun, Tianxi ; Liu, Zhiguo ; Li, Yude ; Wang, Guangfu ; Zhu, Guanghua ; Ding, Xunliang ; Liu, Hui ; Luo, Ping ; Pan, Qiuli ; Lin, Xiaoyan ; Teng, Yuepeng] Beijing Radiat Ctr, Beijing 100875, Peoples R China ; [Xu, Qing] Chinese Acad Sci, Inst High Energy Phys, Beijing 100039, Peoples R China |
英文摘要 | A laboratory micro X-ray fluorescence spectrometer based on polycapillary X-ray optics (PXRO) was used to carry out the quantitative X-ray fluorescence analysis of single aerosol particles with smaller size than that of focal spot of PXRO. The minimum detection limits measured with the thin-film reference standards were in the range from 13.3 to 0.7 ng cm(-2) when the operating current and voltage were 70 mA and 35 kV, respectively. In order to reduce the effects of the inhomogeneous distributions of the X-ray intensity in the focal spot of the PXRO on the analysis results, the sensitivities were corrected by using a Gaussian function for the quantitative analysis of single aerosol particles. The accuracy of the analysis of single standard solution drops was on average 25% depending on the element and concentration. The precision of the analysis was better than 5%. (C) 2009 Elsevier B.V. All rights reserved. |
学科主题 | Spectroscopy |
类目[WOS] | Spectroscopy |
研究领域[WOS] | Spectroscopy |
原文出处 | SCI |
语种 | 英语 |
WOS记录号 | WOS:000272910300005 |
内容类型 | 期刊论文 |
源URL | [http://ir.ihep.ac.cn/handle/311005/239164] |
专题 | 高能物理研究所_多学科研究中心 高能物理研究所_粒子天体物理中心 中国科学院高能物理研究所_中国散裂中子源 |
作者单位 | 中国科学院高能物理研究所 |
推荐引用方式 GB/T 7714 | Sun, TX,Liu, ZG,Li, YD,et al. Quantitative analysis of single aerosol particles using polycapillary X-ray optics[J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY,2009,64(11-12):1194-1197. |
APA | Sun, TX.,Liu, ZG.,Li, YD.,Wang, GF.,Zhu, GH.,...&徐清.(2009).Quantitative analysis of single aerosol particles using polycapillary X-ray optics.SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY,64(11-12),1194-1197. |