Design of the Online Integral Dose Monitor System for BESIII EMC
Gong, H; Hou, L; Gong, GH; Shao, BB; Li, YX; Li, J; Li J(李金)
刊名JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY
2008
页码#REF!
关键词BESIII electro-magnetic calorimeter dose monitor RadFET
通讯作者Gong, H (reprint author), Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China.
英文摘要High beam current of BEPCII leads to high radiation background around the interaction point where the Electro-Magnetic Calorimeter (EMC), a key sub-detector of BESIII, is located. The performance of EMC degrades due to accumulated exposure to the high radiation background. In order to monitor the integral dose deposited in the EMC, an online integral dose monitor system has been developed according to the study on 400nm IMPL RadFET. This system consists of 72 RadFETs located around EMC. A microcontroller based DAQ box can readout all the RadFET information and send to Computers via an Ethernet interface. The detail of this system and the BEPC testing result are presented in this paper.
学科主题Nuclear Science & Technology
类目[WOS]Nuclear Science & Technology
收录类别SCI
WOS记录号WOS:000267196000065
公开日期2016-05-03
内容类型期刊论文
源URL[http://ir.ihep.ac.cn/handle/311005/227218]  
专题高能物理研究所_实验物理中心
推荐引用方式
GB/T 7714
Gong, H,Hou, L,Gong, GH,et al. Design of the Online Integral Dose Monitor System for BESIII EMC[J]. JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY,2008:#REF!.
APA Gong, H.,Hou, L.,Gong, GH.,Shao, BB.,Li, YX.,...&李金.(2008).Design of the Online Integral Dose Monitor System for BESIII EMC.JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY,#REF!.
MLA Gong, H,et al."Design of the Online Integral Dose Monitor System for BESIII EMC".JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY (2008):#REF!.