Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system | |
Wang YF(王燕芳)![]() ![]() ![]() | |
刊名 | CHINESE PHYSICS C
![]() |
2013 | |
卷号 | 37期号:7页码:78202 |
关键词 | micro computed tomography point spread function contrast discrimination function modulation transfer function |
英文摘要 | The spatial resolution and the relative density resolution are the two most critical indicators in CT system. The method recommended in the ASTM E1695-95 and GJB 5311-2004 is only suitable to the fan-beam CT system. In this paper, for industrial cone-beam micro CT system, we will adopt the edge response function (ERF) created by the step edges of a steel ball to measure the system 3D PSF and MTF. To describe the contrast discrimination function more accurately, we will first propose to extend the two-dimensional measurement region to the three-dimensional space. Our experimental spatial resolution is (55.56+/-0.56) lp/mm and the relative density resolution is 1% within 300 mu mx300 mu mx300 mu m according to the 3 sigma rule. |
学科主题 | Physics |
收录类别 | SCI |
WOS记录号 | WOS:000322937400017 |
公开日期 | 2016-05-03 |
内容类型 | 期刊论文 |
源URL | [http://ir.ihep.ac.cn/handle/311005/224759] ![]() |
专题 | 高能物理研究所_核技术应用研究中心 |
推荐引用方式 GB/T 7714 | Wang YF,Que JM,Cao DQ,et al. Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system[J]. CHINESE PHYSICS C,2013,37(7):78202. |
APA | 王燕芳.,阙介民.,曹大泉.,Wang, YF.,Que, JM.,...&魏龙.(2013).Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system.CHINESE PHYSICS C,37(7),78202. |
MLA | 王燕芳,et al."Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system".CHINESE PHYSICS C 37.7(2013):78202. |